CONTROLLER: C9536-01/C9536-02
SENSOR HEAD: H9958 SERIES
UV POWER METERS
C9536 / H9958 SERIES
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications aresubject to change without notice. No patent rights are granted to any of the circuits described herein. ©2011 Hamamatsu Photonics K.K.
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
SPECIFICATIONS
OVER VIEW
FEATURES APPLICATIONS
The C9536/H9958 series are optical power meters designed to measure absolute optical power of UV light (mW/cm2). These optical power meters have a flat spectral response over a wide UV wave-length region, and are capable of measuring the optical power inde-pendent of the spectral emission distribution of a light source to be measured. Unlike commonly available UV power meters, optical pow-er of various UV light sources can be easily measured with just one unit of this power meter, without having correct each wavelength of emitted light. A combination of the C9536-02 and H9958-02 is suit-able for measuring the optical power of high output UV-LED.Also, we recommend a combination of the C9536-01 and H9958 as a power meter to measure the optical power of light sources for photo-catalysis.
Parameter Description / Value
Measurement Mode
External InterfacePower RequirementDimensions (W × H × D)Operating Ambient Temperature / HumidityStorage Temperature / HumidityWeight
This mode makes continuous measurements at a sampling rate of 1 second.This mode displays a result integrated over a certain period of time
(adjustable from 1 second to 5 minutes in 1 second steps). This mode makes one measurement each time START switch is pressed.
This mode holds the peak values in continuous measurement.RS-232C
Internal battery or AC adapter75 mm × 110 mm × 20 mm
0 °C to +40 °C / Below 80 %-20 °C to +50 °C / Below 80 %
Approx. 285 g
Continuous
Integration
One Shot (C9536-01 only)Peak Hold (C9536-02 only)
Parameter Description / ValueController and Sensor Head CombinationTypeSpectral ResponseEffective AreaMeasurement RangeDimensions (W × H × D)Operating Ambient Temperature / HumidityStorage Temperature / HumidityWeight
C9536-01
10 mm1 µW/cm2 to 100 mW/cm2
0 °C to +45 °C / Below 80 %-20 °C to +50 °C / Below 80 %
H9958310 nm to 380 nm
H9958-01
55 mm × 75 mm × 18 mm
Approx. 195 g
C9536-02H9958-02
1 mm100 µW/cm2 to 10 W/cm2
0 °C to +60 °C / Below 80 %-20 °C to +60 °C / Below 80 %
300 nm to 410 nm
Left: Controller C9536-01Right: Sensor Head H9958/-01
Left: Controller C9536-02Right: Sensor Head H9958-02
�Traceable to National Standard in Japan and a Certificate of Calibration is Appended
�Flat Spectral Response Characteristics Over a Wide UV Region�High Sensitivity (detectable down to 1 µW/cm2)�Compatible with high output UV-LED�Ideal for Monitoring / Controlling Light Source Power for Photocatalyst
Monitoring, Measurement and Control of:�Black Light�UV LED�Ultraviolet Rays (UV-A)�Mercury-Xenon Lamp�High Pressure Mercury-Xenon Lamp�Deuterium Lamp
TPT 1026E01FEB. 2011 IP
Figure 4: Dimensional Outline (Unit: mm)
Figure 1: Relative Spectral Response Characteristics of Detector (Typ.) (H9958: 310 nm to 380 nm)
Figure 2: Relative Spectral Response Characteristics of Detector (Typ.) (H9958-01/-02: 300 nm to 410 nm)
TPT B0026EC
Figure 3: Detector Sensitivity vs. Incident Light Angle (Typ.)TPT B0028EC
TPT B0040EB
TPT A0041EB
TPT A0042EB
�Sensor Head: H9958 Series �Controller: C9536 Series
�Controller and Sensor Head Combination
RE
LAT
IVE
VA
LUE
(%
)
20
40
60
80
100
200 250 300 350
WAVELENGTH (nm)
400 450 500
BLACK LIGHT
BLACK LIGHT
H9958
351 368
HAMAMATSU PHOTONICS K.K.HAMAMATSU PHOTONICS K.K., Electron Tube Division 314-5, Shimokanzo, Iwata City, Shizuoka Pref., 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: [emailprotected]: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: [emailprotected]: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: [emailprotected] Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: [emailprotected] Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: [emailprotected]: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: [emailprotected]
www.hamamatsu.com
UV POWER METER C9536, H9958 SERIESR
ELA
TIV
E V
ALU
E (
%)
20
40
60
80
100
ANGLE (degree)
-100 -80 -60 -40 -20 0 20 40 60 80 100
COS θ
H9958/-01
H9958-02
471518
55
26
557520
00
0.5
EFFECTIVE AREA 1
INPUT WINDOW 14EFFECTIVE AREA 10
55
26
7520
00
4-M3 THREDED HOLE
FRONT VIEW(H9958/-01)
FRONT VIEW(H9958-02)
SIDE VIEW SIDE VIEW
SELECT ENTER
UV POWER METER
STARTSTOP POWER
75
110
20
ANALOG OUTPUT CONNECTOR
RS-232C CONNECTOR DC INLET
DC INANALOG OUTRS-232C
LCD
RE
LAT
IVE
VA
LUE
(%
)
20
40
60
80
100
365 385
WAVELENGTH (nm)
200 250 300 350 400 450 500
H9958-01/-02
UV LEDUV LED
Note: Slight response may occur at some wavelength over 500 nm.Note: Slight response may occur at some wavelength over 500 nm.
ControllerC9536-01C9536-02
Sensor HeadH9958, H9958-01H9958-02